GIMA MT Base Plus

GIMA MT Base Plus Magnetotherapy Device User Manual

Model: MT Base Plus (28321)

1. Introduzione

The GIMA MT Base Plus is a low-frequency pulsed magnetotherapy device designed for home use. It generates an impulsive magnetic field.

The device incorporates a square-wave electric current generator with a variable frequency from 10 to 99 Hz and variable pulse duration, allowing for a pre-determinable field intensity from 10 to 80 Gauss.

It also includes an audible magnetic field detector and a safety system that blocks power delivery in case of a short circuit.

Therapeutic Effects

The primary effects of low-frequency magnetic fields include:

  • Anti-edematous (reduces swelling)
  • Analgesic (pain relief)
  • Antinfiammatorio
  • Rigenerativo

These effects are primarily applied in traumatology, rheumatology, angiology, neurology, dermatology, and endocrinology.

2. Informazioni importanti sulla sicurezza

Please read this manual carefully before using the GIMA MT Base Plus device. Failure to follow instructions may result in injury or device malfunction.

Precauzioni generali di sicurezza

  • Utilizzare il dispositivo solo come descritto in questo manuale.
  • Tenere il dispositivo lontano dall'acqua e da altri liquidi.
  • Non utilizzare il dispositivo se è danneggiato.
  • Consult a medical professional before use, especially if you have pre-existing medical conditions, pacemakers, or are pregnant.
  • Tenere fuori dalla portata dei bambini.
  • Non tentare di aprire o riparare il dispositivo da soli. Contattare personale di assistenza qualificato.
  • Ensure the power supply matches the device's requirements (230V/50-60 Hz).

Compatibilità elettromagnetica (EMC)

This device complies with EN60601-1 (CE 162-5) safety standards. It is designed to minimize electromagnetic interference. However, avoid using it near other electronic devices that may cause or be susceptible to interference.

3. Contenuto della confezione

Carefully unpack the device and check that all items are present and undamaged. If any items are missing or damaged, contact your supplier immediately.

  • GIMA MT Base Plus Device
  • Cavo di alimentazione
  • 2 Application Plates (70 x 120 cm)
  • 1 Velcro Elastic Band (60 x 8 cm)
  • 1 Velcro Elastic Band (100 x 8 cm)
  • ABS Carrying Case
  • Manuale dell'utente (questo documento)
GIMA MT Base Plus Magnetotherapy Device and its accessories

Image: The GIMA MT Base Plus device with its included accessories, such as application plates and elastic bands, neatly arranged.

4. Istruzioni per l'installazione

Ispezione iniziale

Before first use, inspect the device and all accessories for any visible damage.

Collegamento dell'alimentazione

  1. Place the GIMA MT Base Plus device on a stable, flat surface.
  2. Assicurarsi che il dispositivo sia spento.
  3. Collegare il cavo di alimentazione all'ingresso di alimentazione del dispositivo.
  4. Plug the other end of the power cable into a standard 230V/50-60 Hz electrical outlet.

Attaching Application Plates

  1. Identify the two application plates. These are the components that deliver the magnetic field.
  2. Connect the application plates to the designated ports on the GIMA MT Base Plus device. Ensure a secure connection.
  3. Use the provided Velcro elastic bands to comfortably secure the application plates to the desired treatment area on the body. Do not overtighten.

5. Utilizzo del dispositivo

Accensione/spegnimento

  1. To power on the device, press the power button. The display should illuminate.
  2. Per spegnere il dispositivo, tenere premuto il pulsante di accensione finché il display non si spegne.

Setting Treatment Parameters

The GIMA MT Base Plus allows adjustment of magnetic field frequency and intensity.

  1. Regolazione della frequenza: Use the frequency control buttons (e.g., '+' and '-') to select the desired frequency between 10 and 99 Hz.
  2. Regolazione dell'intensità: Use the intensity control buttons to set the magnetic field intensity between 10 and 80 Gauss.
  3. Durata dell'impulso: The pulse duration is automatically adjusted based on the selected frequency and intensity to achieve the desired field.

Avvio di una sessione di trattamento

  1. After setting the desired frequency and intensity, press the 'Start' or 'Run' button to begin the treatment.
  2. The device will emit an audible signal if the magnetic field is detected, confirming operation.
  3. Monitor the treatment area and ensure comfort. If any discomfort occurs, stop the treatment immediately.

Rilevatore di campo magnetico

The device includes an audible magnetic field detector. This feature provides confirmation that the magnetic field is being generated. If no sound is heard during operation, check connections and settings.

6. Manutenzione e cura

Pulizia

Scollegare sempre il dispositivo dalla rete elettrica prima della pulizia.

  • Wipe the device's exterior with a soft, damp stoffa.
  • Non utilizzare detergenti abrasivi, solventi o prodotti chimici aggressivi.
  • Assicurarsi che nessun liquido penetri nel dispositivo.
  • Clean application plates with a mild disinfectant solution and allow them to air dry completely.

Magazzinaggio

  • Store the device and its accessories in the provided ABS carrying case when not in use.
  • Conservare in un luogo fresco e asciutto, lontano dalla luce solare diretta e da temperature estreme.
  • Tenere fuori dalla portata dei bambini.

Disposizione

This product should not be disposed of with household waste. Please follow local regulations for the disposal of electronic devices.

7. Risoluzione Dei Problemi

ProblemaPossibile causaSoluzione
Il dispositivo non si accende.Cavo di alimentazione non collegato o presa difettosa.Controllare il collegamento del cavo di alimentazione. Provare una presa diversa.
No magnetic field detected (no audible signal).Application plates not connected properly or device malfunction.Ensure application plates are securely connected. If problem persists, contact support.
Il dispositivo si arresta durante il funzionamento.Short circuit detected by safety system.Turn off the device, check all connections, and restart. If it recurs, contact support.
Il display non funziona correttamente.Internal error or damage.Power cycle the device. If the issue persists, contact support.

If you encounter any issues not listed here or if the suggested solutions do not resolve the problem, please contact GIMA customer support.

8. Specifiche tecniche

CaratteristicaSpecificazione
Forma d'onda di modulazionePiazza
Maximum Emission Ratio50%
Adjustable Magnetic Field10-80 Gauss
Frequenza del campoSelectable from 10 to 99 Hz
Maximum Field Intensity80 Gauss
Alimentazione elettrica230V/ 50-60 Hz
Dimensioni (L x P x A)Dimensioni: 193 x 138 x 68 mm
Peso1 Kg
Classe di sicurezzaII BF
StandardEN60601 - 1 (CE 162-5)
Numero di parte del produttore28321

9. Garanzia e assistenza clienti

Informazioni sulla garanzia

GIMA products are manufactured to high-quality standards. For specific warranty terms and conditions, please refer to the warranty card included with your purchase or visit the official GIMA websito. Conservare la prova d'acquisto per eventuali reclami in garanzia.

Assistenza clienti

For technical assistance, troubleshooting, or service inquiries, please contact GIMA customer support.

You can find contact information on the official GIMA websito o tramite il tuo rivenditore autorizzato.

When contacting support, please have your device model (MT Base Plus) and serial number (if applicable) ready.

Ask a question about this manual