1. Introduzione
This manual provides essential information for the safe and effective operation, setup, and maintenance of your BMC RESmart GII Auto CPAP System E-20A-H with Humidifier. Please read this manual thoroughly before using the device and retain it for future reference. This device is intended for the treatment of Obstructive Sleep Apnea (OSA).
2. Informazioni sulla sicurezza
Avvertenze di sicurezza importanti:
- Always disconnect the device from the power supply before cleaning or performing any maintenance.
- Non immergere il dispositivo in acqua o in altri liquidi.
- Use only accessories and parts recommended by BMC.
- Keep the device away from direct sunlight, heat sources, and sharp objects.
- Consult your physician or healthcare provider if you experience any unusual symptoms or discomfort during use.
- Questo dispositivo non è destinato al supporto vitale.
3. Prodotto finitoview
The BMC RESmart GII Auto CPAP System E-20A-H is an intelligent device designed for the treatment of Obstructive Sleep Apnea. It features an integrated humidifier for enhanced comfort.
Componenti chiave:
- Unità principale: Houses the motor, control electronics, and display.
- Umidificatore: Integrated unit with dual water chambers for efficient humidification.
- Display LCD da 3.5 pollici: Provides real-time information and access to settings.
- Manopola di controllo: Per navigare nei menu e regolare le impostazioni.
- Uscita aria: Connects to the CPAP tubing.
- Ingresso di alimentazione: Per collegare l'adattatore di alimentazione.
- Slot per scheda SD: For compliance data storage and retrieval (iCode II).
- SpO2 Kit Port (Optional): For real-time blood oxygen saturation and pulse rate monitoring.

An image showing the BMC RESmart GII Auto CPAP device, featuring its main unit, control knob, and integrated humidifier. The device is black with a clear humidifier chamber visible.
4. Impostazione
4.1 Disimballaggio
- Rimuovere con attenzione tutti i componenti dalla confezione.
- Ispezionare il dispositivo e gli accessori per verificare la presenza di eventuali segni di danneggiamento. Non utilizzare se danneggiati.
4.2 Collegamento dell'alimentazione
- Connect the power adapter to the power inlet on the back of the device.
- Collegare il cavo di alimentazione a una presa elettrica idonea.
4.3 Filling the Humidifier
- Open the humidifier chamber by sliding or lifting the cover.
- Fill the water chamber with distilled water up to the maximum fill line. Do not overfill.
- Close the humidifier chamber securely.
4.4 Connecting Tubing and Mask
- Connect one end of the CPAP tubing to the air outlet on the device.
- Connect the other end of the tubing to your prescribed CPAP mask.
- Fit the mask comfortably and securely according to its own instructions.
5. Istruzioni per l'uso
5.1 Inizio della terapia
- Ensure the device is properly set up and connected.
- Indossa la mascherina.
- Press the Start/Stop button (usually the control knob) to begin therapy. The device will automatically start delivering air.
5.2 Regolazione delle impostazioni
Use the control knob to navigate the menu on the 3.5-inch LCD display. Press the knob to select an option and rotate to adjust values.
- Intervallo di pressione: 4-20 cm H2O (Auto CPAP adjusts automatically within this range).
- Ramp Tempo: Adjusts the time (0-60 minutes) for the pressure to gradually increase to the prescribed level.
- Humidification Level: Select from Off or levels 1-5 to control the humidity output.
- Reslex: Pressure relief feature (Off/1/2/3) for exhalation comfort.
- Accensione/spegnimento automatico: Can be enabled or disabled (Off by default).
5.3 Interruzione della terapia
Press the Start/Stop button again to end therapy. The device will power down.
5.4 iCode II Data Retrieval
Compliance data is encoded by a QR code displayed on the device's screen. Use a compatible application to scan the QR code for quick data retrieval and monitoring.
6. Manutenzione
Regular cleaning and maintenance are crucial for the optimal performance and longevity of your device.
6.1 Pulizia giornaliera
- Humidifier Water Chamber: Empty any remaining water and rinse with distilled water. Allow to air dry.
- Maschera: Wipe down the mask cushion with a damp cloth or specialized CPAP wipes.
6.2 Pulizia settimanale
- Humidifier Water Chamber: Wash with mild soap and warm water. Rinse thoroughly and allow to air dry completely.
- Tubi CPAP: Wash with mild soap and warm water. Rinse thoroughly and hang to air dry.
- Filtro dell'aria: Inspect the reusable air filter (if applicable). If dirty, rinse under running water and allow to air dry. Replace disposable filters as recommended.
6.3 Cura generale
- Pulisci l'esterno del dispositivo con adamp cloth. Do not use abrasive cleaners or immerse the device.
- Replace air filters regularly as per manufacturer recommendations or if visibly dirty.
7. Risoluzione Dei Problemi
In caso di problemi, fare riferimento ai seguenti problemi comuni e alle relative soluzioni:
| Problema | Possibile causa | Soluzione |
|---|---|---|
| Nessun potere | Cavo di alimentazione scollegato; Alimentazione spentatage | Check power connections; Restore power supply |
| Air leak from mask | Mask not fitted correctly; Mask worn out | Adjust mask straps; Replace mask cushion or mask |
| Secchezza della bocca/gola | Humidifier off or set too low; Water chamber empty | Increase humidification level; Refill water chamber |
| Il dispositivo è rumoroso | Air filter dirty or blocked; Device on an uneven surface | Clean or replace air filter; Place device on a stable, flat surface |
| Pressure feels too high/low | Incorrect pressure setting; Ramp feature active | Consult your physician for pressure adjustments; Wait for ramp per completare |
If the problem persists, contact your healthcare provider or the manufacturer's customer support.
8. Specifiche
| Caratteristica | Specificazione |
|---|---|
| Modello | E-20A-H |
| Modalità | CPAP / Auto |
| Campo di pressione | 4-20 cmH2O |
| Pressure Increment | 0.5 cm H2O |
| Ramp Tempo | 0-60 minuti |
| Rislex | Spento/1/2/3 |
| Livello sonoro | <30 dB |
| Humidification Levels | Spento / 1-5 |
| Display | LCD da 3.5 pollici |
| Dimensioni (LxWxH) | 170x196x118mm (device) / 290x196x134mm (with H60 humidifier) |
| Peso | <1.5kg (device) / <2.5kg (with H60 humidifier) |
| Alimentazione elettrica | Input Range 100-240V AC, 50/60 Hz, 2.0 A max |
| Archiviazione dei dati | SD Card (iCode II) |
| Paese di origine | Cina |
9. Garanzia e supporto
For warranty information, please refer to the documentation provided with your purchase or contact your retailer. For technical support, service, or questions regarding your BMC RESmart GII Auto CPAP System, please contact the seller or the manufacturer directly.