VTSYIQI ETCR007

VTSYIQI ETCR007 AC/DC Clamp Leakage Current Sensor User Manual

Model: ETCR007 | Brand: VTSYIQI

1. Introduzione

The VTSYIQI ETCR007 AC/DC Clamp Leakage Current Sensor is designed for precise measurement of AC/DC current and leakage current. It operates under line voltages up to 600V. This device utilizes advanced CT technology and digital integration for accurate, non-contact current measurement. It is suitable for various applications including power analysis, electrical testing, industrial maintenance, and scientific research.

This manual provides essential information for the safe and effective use of your ETCR007 sensor. Please read it thoroughly before operation.

2. Caratteristiche del prodotto

  • Funzione: Measures AC/DC current and leakage current.
  • Metodo di misurazione: Clip-on CT, non-contact measurement.
  • Clamp Misurare: φ7 mm.
  • Intervallo di input: 0mA to 60.0A AC/DC.
  • Volume di uscitatage: Selectable 10mV/A or 100mV/A.
  • Risoluzione: 1mA AC/DC.
  • Precisione: ±3%FS (at 23°C ±5°C, below 5%RH).
  • Errore di fase: ≤3° (AC 50Hz/60Hz; 23°C ±2°C).
  • Gamma di frequenza: AC: 45Hz-400Hz; DC: DC-100kHz.
  • Interfaccia di uscita: BNC plug.
  • Alimentazione elettrica: 9V battery (6LR61/6F22) or external power supply.
  • Standard di sicurezza: Complies with IEC1010-1, IEC1010-2-032, Pollution level 2, CAT ΙΙΙ (600V).

3. Informazioni sulla sicurezza

To ensure safe operation and prevent electric shock or damage to the device, please observe the following safety precautions:

  • Always disconnect the line being tested before opening the clamp sensor housing.
  • Do not use the sensor on circuits exceeding AC600V line voltage.
  • Assicurarsi che il vano batteria sia ben chiuso durante il funzionamento.
  • Do not operate the device in wet environments or if it appears damaged.
  • Refer to the applicable safety rules: IEC1010-1, IEC1010-2-032, Pollution level 2, CAT ΙΙΙ (600V).
  • Keep the device away from strong electromagnetic fields to avoid measurement interference.

4. Contenuto della confezione

Verificare che tutti gli articoli siano presenti e integri al momento del disimballaggio:

  • AC/DC Clamp Leakage Current Sensor (1 piece)
  • 9V Battery (6F22, 1 piece)
  • User Manual / Warranty Card / Certification (1 copy)
VTSYIQI ETCR007 AC/DC Clamp Leakage Current Sensor and accessories in box

Image 1: The VTSYIQI ETCR007 clamp sensor, output cable, and user manual, shown partially in its packaging box.

VTSYIQI ETCR007 AC/DC Clamp Leakage Current Sensor, cable, and manual removed from box

Image 2: The VTSYIQI ETCR007 clamp sensor, its attached cable, and the user manual, displayed outside of the product box.

Video 1: An unboxing video demonstrating the contents of the VTSYIQI ETCR007 package, including the clamp sensor and accessories.

5. Impostazione

5.1 Installazione della batteria

  1. Individuare il coperchio del vano batteria sul retro del sensore.
  2. Open the cover and insert the provided 9V (6F22) battery, ensuring correct polarity.
  3. Chiudere saldamente il coperchio del vano batterie.

5.2 Connecting the Output Wire

The sensor comes with an integrated output cable. The core wire provides the signal positive output, and the shielding wire serves as the signal ground. Connect the BNC plug of this cable to a compatible measuring instrument (e.g., digital multimeter, oscilloscope, power quality analyzer).

Close-up of the BNC connector on the VTSYIQI ETCR007 sensor's output cable

Immagine 3: Un dettaglio view of the BNC connector at the end of the sensor's output cable, ready for connection to a measuring device.

6. Funzionamento

6.1 Powering On and Range Selection

  1. Locate the POWER switch on the sensor.
  2. Slide the switch to select the desired output range: 10mV / A or 100mV / A. The "OFF" position powers down the device.
  3. The selected range determines the output voltage per ampere of measured current. For example, 100mV/A means 100mV output for every 1A of current.
Close-up of the power switch and output range selector on the VTSYIQI ETCR007 sensor

Immagine 4: Un primo piano view of the sensor's control panel, showing the power switch with 10mV/A, 100mV/A, and OFF positions.

6.2 Calibrazione zero

Before taking measurements, especially for DC current or leakage current, it is crucial to perform a zero calibration to eliminate external electric field interference and ensure accuracy.

  1. Assicurarsi che non vi sia corrente che scorre attraverso il clamp mascella.
  2. Locate the ZERO adjustment knob (often a rotary knob) on the sensor.
  3. Rotate the knob until the connected measuring instrument displays zero current.
Primo piano del clamp jaw and zero adjustment area on the VTSYIQI ETCR007 sensor

Immagine 5: Un dettaglio view del clamp jaw and the area where the zero adjustment knob is typically located, indicated by a yellow arrow symbol.

6.3 Esecuzione delle misurazioni

  1. Aprire il clamp mascella premendo il grilletto.
  2. Place the wire carrying the current to be measured inside the center of the clamp jaw. Ensure only one conductor is within the jaw for accurate measurement.
  3. Rilasciare il grilletto per chiudere la clamp stringere saldamente la ganascia attorno al conduttore.
  4. Read the current value from your connected measuring instrument, applying the selected output ratio (e.g., if 100mV/A is selected and the instrument reads 500mV, the current is 5A).
  5. For optimal accuracy, ensure the tested wire is positioned in the jaw's center.
Lato view of the VTSYIQI ETCR007 AC/DC Clamp Leakage Current Sensor

Immagine 6: Prospettiva lateralefile of the VTSYIQI ETCR007 clamp sensor, showing the ergonomic design for gripping and clamping around a conductor.

7. Specifiche

ParametroValore
FunzioneMeasurement of AC/DC current, leakage current
Alimentazione elettrica9V Battery (6LR61/6F22), or external power supply
Durata della batteriaCirca 100 ore di utilizzo continuo
Modalità di provaClip-on CT, non-contact measurement
Clamp Misurareφ7mm
Intervallo di input0mA - 60.0A AC/DC
Volume di uscitatage10mV/A; 100mV/A (corresponding)
Gamma di uscita1V peak max
Risoluzione1mA AC / DC
Precisione±3%FS (23°C ±5°C, below 5%RH)
Errore di fase≤3° (AC 50Hz/60Hz; 23°C ±2°C)
CalibrazioneAdjust ZERO to calibrate
Lunghezza del filo di uscita2m
Dimensioni (L x P x A)168 mm × 65 mm × 34 mm
Frequenza (CA)45Hz - 400Hz
Frequency (DC)DC - 100kHz
Posizione di provaTested wire in the jaw center
Linea VoltageAC600V
Peso170 g (batteria inclusa)
Corrente di lavoro5mA
Temperatura di lavoro-10°C a 50°C
Umidità di lavoroSotto 80% RH
Temperatura di conservazione-10°C a 60°C
Umidità di stoccaggioSotto 70% RH
Forza isolanteAC3700V/rms (between alloy of clamp and housing)
Applicable Safety RulesIEC1010-1, IEC1010-2-032, Pollution level 2, CAT ΙΙΙ (600V)
Posteriore view of the VTSYIQI ETCR007 sensor showing product label with barcode 00770112 and specifications

Image 7: The back of the sensor, displaying a product label with model information, specifications, and a barcode (00770112).

8. Manutenzione

  • Pulizia: Use a soft, dry cloth to clean the sensor. Do not use abrasive cleaners or solvents.
  • Magazzinaggio: When not in use for extended periods, remove the battery to prevent leakage. Store the device in a dry environment, away from direct sunlight and extreme temperatures.
  • Sostituzione della batteria: Replace the 9V battery when the power indicator light dims or the device fails to power on.
  • Calibrazione: Regular calibration by a qualified technician is recommended to maintain measurement accuracy, especially after prolonged use or if accuracy is suspected to have drifted.

9. Risoluzione Dei Problemi

ProblemaPossibile causaSoluzione
Il dispositivo non si accendeBatteria scarica o installata in modo erratoControllare la polarità della batteria e sostituirla con una nuova batteria da 9 V.
Letture imprecise
  • - Improper zero calibration
  • - Tested wire not centered in jaw
  • - External electromagnetic interference
  • - Multiple conductors in jaw
  • - Perform zero calibration before measurement.
  • - Ensure wire is centered.
  • - Move away from strong magnetic fields.
  • - Ensure only one conductor is clamped.
Nessun segnale di uscita
  • - Loose or faulty output cable connection
  • - Measuring instrument not configured correctly
  • - Check cable connection to both sensor and measuring instrument.
  • - Verify measuring instrument settings (e.g., input type, range).

10. Garanzia e supporto

This VTSYIQI ETCR007 AC/DC Clamp Leakage Current Sensor comes with a Garanzia di 1 anno dalla data di acquisto, coprendo i difetti di materiali e di fabbricazione in condizioni di normale utilizzo.

For technical support, warranty claims, or service inquiries, please contact the manufacturer:

Si prega di conservare la ricevuta d'acquisto come prova d'acquisto ai fini della garanzia.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.