MiLESEEY MC996

MiLESEEY Digital Coating Thickness Gauge MC996 User Manual

1. Introduzione

The MiLESEEY MC996 Digital Coating Thickness Gauge is designed for precise measurement of non-magnetic coatings on magnetic metal substrates (ferrous, Fe) and non-conductive coatings on non-ferrous metal substrates (NF). This device features automatic F/NF mode detection, fast readings, zero and multi-point calibration, unit conversion (metric/imperial), single and continuous measurement modes, and an automatic shut-down function for energy saving. It is suitable for various applications including automotive, manufacturing, and quality inspection.

2. Contenuto della confezione

Verificare che tutti gli articoli siano presenti nel pacco:

  • 1 x MiLESEEY Coating Thickness Gauge (MC996)
  • 1 x Aluminium Substrate
  • 1 x substrato di ferro
  • 5 x Standard Calibration Foils
  • 2 batterie AAA
  • 1 x Manuale utente
  • 1 x Toolbox/Carrying Case
MiLESEEY MC996 package contents including gauge, calibration foils, and substrates

Figure 2.1: Included accessories for the MiLESEEY MC996 gauge, showing calibration foils, aluminum and iron substrates, and the carrying case.

3. Prodotto finitoview

Familiarize yourself with the components and display indicators of the MC996 gauge.

Labeled diagram of MiLESEEY MC996 gauge showing screen, buttons, and sensor

Figure 3.1: Labeled diagram of the MiLESEEY MC996 Digital Coating Thickness Gauge, highlighting the 2.2-inch backlit screen, units switch, calibration buttons, mode switch, power button, and detecting sensor.

Componenti chiave:

  • Schermo retroilluminato da 2.2": Displays measurement readings, units, battery status, and mode indicators.
  • CALIB/ZERO Button: Used for zero calibration and reducing calibration data.
  • um/mils Button: Switches between metric (µm) and imperial (mils) units, and increases calibration data.
  • MODE/RESET Button: Switches between Single (SNG) and Continuous (CON) measurement modes, and resets to default settings.
  • Pulsante di accensione: Accende e spegne il dispositivo.
  • Detecting Sensor: Located at the bottom, used to contact the surface for measurement.

4. Impostazione

4.1 Installazione della batteria

  1. Individuare il vano batteria sul retro del dispositivo.
  2. Aprire il coperchio della batteria.
  3. Inserire due batterie AAA, assicurandosi della corretta polarità (+/-).
  4. Chiudere bene il coperchio della batteria.

4.2 Accensione iniziale

Press the power button to turn on the device. The screen will light up, and the gauge will perform a self-check. Ensure the sensor is not touching any surface during power-on.

5. Istruzioni per l'uso

5.1 Modalità di misurazione

The gauge supports two measurement modes:

  • SNG (Single Measurement): Takes one reading each time the sensor is placed on the surface.
  • CON (Continuous Measurement): Continuously takes readings as the sensor is moved across the surface.

Premere il tasto MODALITÀ/RESET button to switch between SNG and CON modes.

5.2 Conversione unità

Premere il tasto um/mil button to switch between micrometers (µm) and mils (mil).

5.3 Automatic F/NF Detection

The MC996 automatically detects the substrate material (Ferrous 'Fe' or Non-Ferrous 'NF') when the sensor is placed on the surface. The detected material type will be displayed on the screen.

MiLESEEY MC996 showing Fe and NF measurement ranges and auto-switch detection

Figure 5.1: The gauge's automatic detection of Ferrous (Fe) and Non-Ferrous (NF) materials, along with their respective measurement ranges and resolutions.

5.4 Esecuzione di una misurazione

  1. Ensure the gauge is powered on and calibrated (see Section 6).
  2. Place the detecting sensor firmly and perpendicularly onto the coated surface to be measured.
  3. The gauge will automatically detect the substrate type and display the coating thickness reading on the screen.
  4. For continuous measurement, slide the sensor across the surface. For single measurement, lift and re-place the sensor for each new reading.
MiLESEEY MC996 'How to Use' steps for detecting car paint thickness

Figura 5.2: Esample of using the MiLESEEY MC996 to detect car paint thickness, including initial zeroing and subsequent measurements.

6. Calibrazione

Calibration is crucial for accurate measurements. The MC996 supports zero calibration and multi-point calibration.

6.1 Calibrazione zero

Zero calibration should be performed on an uncoated substrate of the same material as the one you intend to measure (e.g., bare iron for ferrous measurements, bare aluminum for non-ferrous measurements).

  1. Place the sensor firmly on the bare substrate.
  2. Premere e tenere premuto il tasto CALIB/ZERO button until the display shows "0.00" or a similar indication of zero.
  3. Release the button. The gauge is now zero-calibrated for that substrate type.

6.2 Calibrazione multi-punto

Multi-points calibration uses standard foils of known thickness to ensure accuracy across the measurement range. Use the included calibration foils and substrates.

  1. Perform zero calibration first (Section 6.1).
  2. Place a calibration foil of known thickness onto the appropriate substrate (e.g., 100µm foil on the iron substrate).
  3. Place the sensor on top of the foil. The gauge will display a reading.
  4. Utilizzare il um/mil E CALIB/ZERO buttons to adjust the displayed value to match the known thickness of the foil.
  5. Repeat for other calibration foils to cover the desired measurement range.

7. Manutenzione

7.1 Pulizia

Wipe the device with a soft, dry cloth. Do not use abrasive cleaners or solvents. Keep the sensor clean and free from debris to ensure accurate readings.

7.2 Sostituzione della batteria

When the low battery indicator appears on the screen, replace the AAA batteries as described in Section 4.1.

7.3 Conservazione

Store the gauge in its carrying case in a dry, cool place, away from direct sunlight and extreme temperatures. Remove batteries if the device will not be used for an extended period.

8. Risoluzione Dei Problemi

ProblemaPossibile causaSoluzione
Il dispositivo non si accendeBatterie scariche o installate in modo erratoSostituire le batterie assicurandosi che la polarità sia corretta.
Gauge powers on but does not take measurementsInternal error or sensor issueRemove batteries for 30 seconds, then reinsert to reset the device. If the problem persists, contact support.
Letture impreciseImproper calibration, dirty sensor, or incorrect measurement techniquePerform zero and multi-point calibration (Section 6). Clean the sensor. Ensure the sensor is placed firmly and perpendicularly.
Il display è scuro o tremolanteBatteria scaricaSostituire le batterie.

9. Specifiche

ParametroValore
ModelloMC996
Campo di misura0 to 1500µm (0 to 59mil)
Risoluzione0.1µm (<100µm), 1µm (≥100µm) / 0.01mil
Precisione±(3%+2µm) / ±(3%+0.08mil)
Minimum Thickness of Metal Substrate0.5mm (Ferrous), 0.3mm (Non-Ferrous)
Tipo di materialeaddominali
Alimentazione elettrica2 batterie AAA
Spegnimento automaticoDopo 5 minuti di inattività
Dimensioni (circa)4.8" (altezza) x 2.6" (larghezza) x 1.1" (profondità)
MiLESEEY MC996 dimensions diagram

Figure 9.1: Approximate dimensions of the MiLESEEY MC996 gauge.

10. Garanzia e supporto

The MiLESEEY MC996 Digital Coating Thickness Gauge typically comes with a manufacturer's warranty. Please refer to the warranty card included in your package or the seller's terms of sale for specific details regarding warranty duration and coverage.

For technical assistance or warranty claims, it is recommended to contact the seller or manufacturer directly. Please note that some users have reported difficulties in receiving timely customer support from the manufacturer. It is advisable to retain your proof of purchase for any warranty-related inquiries.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.