AmScope B120C-DK-E1

AmScope B120C-DK-E1 Binocular LED Darkfield Microscope User Manual

Model: B120C-DK-E1

Introduzione

This manual provides instructions for the proper setup, operation, and maintenance of your AmScope B120C-DK-E1 Binocular LED Darkfield Microscope. This instrument is designed for educational, clinical, and laboratory applications, offering both brightfield and darkfield observation capabilities. Please read this manual thoroughly before using the microscope to ensure safe and effective operation.

Informazioni sulla sicurezza

  • Maneggiare sempre il microscopio con cura. Evitare movimenti o urti bruschi.
  • Assicurarsi che il microscopio sia posizionato su una superficie stabile e piana per evitare che si ribalti.
  • Scollegare il cavo di alimentazione prima di effettuare operazioni di pulizia o manutenzione.
  • Non esporre il microscopio alla luce solare diretta, a temperature elevate o a umidità elevata.
  • Utilizzare solo l'adattatore di alimentazione specificato.
  • Keep the optical components clean. Avoid touching lenses with bare hands.
  • Do not attempt to disassemble or repair the microscope yourself. Contact qualified service personnel.

Componenti e caratteristiche

Familiarize yourself with the main components of your AmScope B120C-DK-E1 microscope:

AmScope B120C-DK-E1 Microscope with labeled features

Figura 1: Sopraview of the AmScope B120C-DK-E1 Binocular Compound Microscope, highlighting key features such as the 2-layer mechanical stage, 360-degree adjustable head, 40X-2500X magnification, coaxial coarse and fine focus, and LED sub-stage illuminazione.

Siedentopf Binocular Head with 360-degree rotation

Figura 2: The Siedentopf binocular head, featuring 360-degree rotation and a 30-degree inclination for comfortable viewing and posture adjustment.

Meccanico Stage and Focus Knobs

Figura 3: Dettagliato view of the 2-layer mechanical stage and the coaxial coarse and fine focus knobs, designed for precise specimen movement and focusing.

LED Transmitted Illumination System

Figura 4: The LED sub-stage lighting system, providing cool, energy-efficient, and daylight-balanced illumination for improved contrast and resolution.

Componenti chiave:

  • Siedentopf Binocular Head: Rotates 360 degrees with a 30-degree inclination for ergonomic viewing.
  • Oculari: Includes 10X and 25X widefield eyepieces for various magnification levels.
  • Lenti dell'obiettivo: Achromatic, color-coded objective lenses (e.g., 4X, 10X, 40X, 100X oil) mounted on a revolving nosepiece.
  • Meccanico Stage: A 2-layer stage with low-position controls for precise X-Y movement of slides.
  • Coaxial Focus Knobs: Separate coarse and fine focus adjustments for accurate focusing.
  • Illuminatore a LED: Sottotitolitage LED light source with adjustable brightness.
  • Condensatore: Includes a brightfield condenser and a darkfield condenser for specialized observation.
  • 1MP Digital Eyepiece Camera: For capturing images and videos, compatible with Windows, Mac, and Linux.

Impostare

  1. Disimballaggio: Rimuovere con cautela tutti i componenti dall'imballaggio. Conservare l'imballaggio per eventuali trasporti o rimessaggio futuri.
  2. Posizionamento: Place the microscope on a stable, vibration-free surface away from direct sunlight and heat sources.
  3. Installazione della testa: Loosen the head retaining screw on the top of the microscope body. Gently insert the binocular head into the opening, ensuring it is fully seated. Tighten the retaining screw.
  4. Installazione dell'oculare: Remove the protective caps from the eyepiece tubes. Insert the desired eyepieces (10X or 25X) into the tubes.
  5. Objective Lens Installation: Rotate the revolving nosepiece to ensure there is no objective lens directly above the stage. Screw the objective lenses into the nosepiece in increasing order of magnification (e.g., 4X, 10X, 40X, 100X).
  6. Collegamento elettrico: Collegare il cavo di alimentazione all'ingresso di alimentazione del microscopio e quindi a una presa elettrica adatta.
  7. Digital Eyepiece Camera Installation: If using the digital eyepiece, remove one of the standard eyepieces and insert the 1MP digital eyepiece camera into the eyepiece tube. Connect the camera to your computer via USB.

Istruzioni per l'uso

1. Brightfield Observation

Microscope objective lenses and magnification examples

Figura 5: Examples of magnification levels (10X, 100X, 400X, 1000X) achievable with the AmScope B120C-DK-E1, utilizing various objective lenses and eyepieces, including the additional 25X eyepieces for up to 2500X magnification.

  1. Accensione: Turn on the main power switch. Adjust the LED illuminator brightness using the dimmer control.
  2. Stage Regolazione: Abbassare la meccanica stage using the coarse focus knob to create enough space to place a slide.
  3. Posizionamento della diapositiva: Place a prepared specimen slide onto the mechanical stage, fissandolo con la stage clips. Use the X-Y stage controls to position the specimen over the light aperture.
  4. Selezione dell'ingrandimento: Rotate the revolving nosepiece to select the lowest power objective lens (e.g., 4X). Ensure it clicks into place.
  5. Messa a fuoco:
    • While looking through the eyepieces, slowly raise the stage using the coarse focus knob until the specimen comes into rough focus.
    • Use the fine focus knob to achieve sharp focus.
    • Adjust the interpupillary distance (distance between eyepieces) and diopter adjustment on the eyepiece for comfortable viewing and to correct for differences in vision between your eyes.
  6. Incrementoasing Ingrandimento: Once focused at low power, rotate the nosepiece to a higher power objective (e.g., 10X, 40X). The microscope is parfocal, meaning only minor adjustments with the fine focus knob should be needed.
  7. Using 100X Oil Immersion Objective:
    • After focusing with the 40X objective, rotate the nosepiece halfway between the 40X and 100X objectives.
    • Place a small drop of immersion oil directly onto the center of the specimen slide.
    • Rotate the 100X objective into the oil drop. The objective lens should make contact with the oil.
    • Use only the fine focus knob to bring the image into sharp focus. Do not use the coarse focus knob with the 100X objective.
    • After use, clean the 100X objective lens and the slide immediately with lens cleaning solution and lens paper.
  8. Regolazione del condensatore: Adjust the iris diaphragm on the condenser to control the amount of light and contrast. For higher magnifications, open the diaphragm more. For lower magnifications, close it slightly. The condenser can also be centered for optimal illumination.

2. Darkfield Observation

The AmScope B120C-DK-E1 includes a darkfield condenser for observing unstained, live specimens.

  1. Condenser Exchange: Remove the brightfield condenser from its mount beneath the stage. Carefully insert the darkfield condenser into the mount. Ensure it is seated correctly.
  2. Preparazione del campione: Darkfield observation works best with thin, unstained specimens.
  3. Illuminazione: Turn on the LED illuminator and set it to a high brightness level.
  4. Messa a fuoco: Use a low power objective (e.g., 10X or 20X, if available) first. Focus on the specimen as you would in brightfield. The specimen will appear brightly lit against a dark background.
  5. Centering the Darkfield Condenser: Some darkfield condensers require centering. If your condenser has centering screws, adjust them until the darkfield effect is uniform across the field of view.
  6. Higher Magnification: For higher magnifications, you may need to use a specialized darkfield objective or ensure your standard objectives are compatible. Always ensure sufficient immersion oil is used with 100X darkfield objectives.

3. Digital Eyepiece Camera Operation

The included 1MP digital eyepiece camera allows for image capture and live viewutilizzando un computer.

Microscope connected to laptop with digital image

Figura 6: The AmScope B120C-DK-E1 microscope connected to a laptop, demonstrating live digital viewing and image capture capabilities via the 1MP digital eyepiece camera.

  1. Installazione del software: Install the professional microscopy software provided (or download the latest version from the AmScope website) onto your Windows, Mac, or Linux computer.
  2. Collega la fotocamera: Insert the digital eyepiece camera into one of the microscope's eyepiece tubes and connect it to your computer via the USB cable.
  3. Software di lancio: Open the installed microscopy software. The software should detect the camera automatically.
  4. Vivere View: The software will display a live feed from the microscope. Adjust microscope focus and illumination for the best image.
  5. Acquisizione immagini/video: Use the software's functions to capture still images or record videos of your observations. The software also provides tools for image processing and measurement.

Manutenzione

  • Pulizia delle lenti: Use a soft brush to remove dust from lens surfaces. For smudges or oil, use a small amount of lens cleaning solution on lens paper. Wipe gently in a circular motion. Never use abrasive cloths or harsh chemicals.
  • Pulizia del corpo: Pulisci il corpo del microscopio con un panno morbido eamp panno. Evitare che l'umidità penetri nei componenti elettrici.
  • Magazzinaggio: When not in use, cover the microscope with a dust cover to protect it from dust and debris. Store in a cool, dry place.
  • Olio da immersione: Always clean immersion oil from the 100X objective and slides immediately after use. Dried oil can be difficult to remove and can damage the lens.

Risoluzione dei problemi

Problema Possibile causa Soluzione
Nessuna illuminazione Power cord disconnected, power switch off, dimmer set to minimum, LED bulb failure. Check power connections, turn on switch, adjust dimmer, contact support if LED is faulty.
Immagine poco chiara/sfocata Incorrect focus, objective lens dirty, eyepiece dirty, slide upside down, condenser not adjusted. Adjust coarse/fine focus, clean lenses, ensure slide is correct side up, adjust condenser/diaphragm.
Macchie scure nel campo di view Polvere sull'oculare, sull'obiettivo o sul vetrino. Clean eyepieces, objectives, and slides. Rotate eyepiece to determine if dust is on eyepiece.
No darkfield effect Incorrect condenser, condenser not centered, illumination too low. Ensure darkfield condenser is installed, center the condenser, increase illumination brightness.
Fotocamera digitale non rilevata USB cable disconnected, software not installed, driver issue. Check USB connection, install software/drivers, restart computer.

Specifiche

Modello B120C-DK-E1
Testa Siedentopf Binocular, 360° Rotatable, 30° Inclined
Oculari WF10X, WF25X
Lenti dell'obiettivo Achromatic 4X, 10X, 40X (Spring), 100X (Spring, Oil)
Ingrandimento totale 40X-2500X
Revolver nasale Revolving Quadruple
Stage 2-Layer Mechanical Stage with X-Y Scale
Messa a fuoco Regolazione coassiale grossolana e fine
Illuminazione LED Transmitted Light with Dimmer
Condensatore Abbe NA1.25 (Brightfield), Darkfield Condenser
Macchina fotografica digitale 1MP USB Digital Eyepiece Camera
Fonte di alimentazione Elettrico con cavo
Materiale Metallo
Colore Bianco

Garanzia e supporto

The AmScope B120C-DK-E1 microscope comes with a Garanzia del produttore di 5 anno. For warranty claims, technical support, or service inquiries, please contact AmScope customer service directly.

For additional resources, software downloads, and frequently asked questions, please visit the official AmScope websito:

www.amscope.com