Model: SM-CAL1

1. Introduzione

IL Amprobe SM-CAL1 is a precision sound meter calibrator designed to ensure the accuracy and reliability of your sound level measurements. This device generates two fixed sound pressure levels, 94 decibels (dB) and 114 dB, at a frequency of 1,000 hertz (Hz), allowing for precise calibration of sound meters with microphones up to 1/2 inch in diameter. Regular calibration is crucial for maintaining measurement integrity and compliance with relevant standards.

This manual provides detailed instructions for the proper use, maintenance, and troubleshooting of your SM-CAL1 Sound Meter Calibrator.

2. Informazioni sulla sicurezza

Please read and understand all safety instructions before operating the SM-CAL1. Failure to follow these guidelines may result in injury or damage to the device.

  • Do not expose the calibrator to extreme temperatures, humidity, or direct sunlight.
  • Evitare di far cadere o sottoporre il dispositivo a forti impatti.
  • Do not attempt to open or modify the calibrator. Refer all servicing to qualified personnel.
  • Assicurarsi che il vano batteria sia ben chiuso durante il funzionamento.
  • Smaltire le batterie secondo le normative locali.

3. Contenuto della confezione

Verificare che tutti gli articoli siano presenti nel pacco:

  • Amprobe SM-CAL1 Sound Meter Calibrator
  • One (1) 9V battery (installed)
  • Istruzioni per l'utente (questo manuale)

4. Impostazione

4.1 Installazione della batteria

The SM-CAL1 comes with a 9V battery pre-installed. If replacement is needed:

  1. Individuare il coperchio del vano batterie nella parte inferiore dell'unità.
  2. Togliere il coperchio.
  3. Connect a new 9V battery, observing correct polarity.
  4. Riposizionare saldamente il coperchio del vano batteria.

4.2 Preparing for Calibration

Before calibrating your sound meter, ensure its microphone is clean and free of debris. The SM-CAL1 is designed to fit microphones with a diameter of up to 1/2 inch.

Amprobe SM-CAL1 Sound Meter Calibrator with power switch and display showing 94 114 OFF db at 1kHz, IEC 942 1988 CLASS 2, and Sound Calibrator.

Figura 1: Amprobe SM-CAL1 Sound Meter Calibrator. The image shows the top face of the cylindrical calibrator, featuring the 'AMPROBE' logo, 'SM-CAL1' model number, a three-position slide switch labeled '94', 'OFF', and '114' (referring to dB levels), and text indicating 'db at 1kHz', 'IEC 942 1988 CLASS 2', and 'Sound Calibrator'. A red power indicator light is also visible.

5. Funzionamento

5.1 Calibrating a Sound Meter

  1. Gently insert the microphone of your sound level meter into the opening on the top of the SM-CAL1 calibrator. Ensure a snug fit to minimize sound leakage.
  2. Select the desired calibration level using the slide switch on the calibrator:
    • Slide to '94' for a 94 dB signal.
    • Slide to '114' for a 114 dB signal.
  3. Observe the reading on your sound level meter. Adjust your sound meter's calibration settings until its display matches the selected output level of the SM-CAL1 (either 94 dB or 114 dB). Refer to your sound meter's user manual for specific calibration adjustment procedures.
  4. Once calibration is complete, slide the switch to 'OFF' to conserve battery life.
  5. Carefully remove the sound meter's microphone from the calibrator.

5.2 Indicazione di batteria scarica

The SM-CAL1 is equipped with a low-battery indicator. When the battery voltage drops below a certain level, the indicator light will dim or not illuminate, signaling that the 9V battery needs to be replaced.

6. Manutenzione

6.1 Pulizia

To clean the exterior of the calibrator, use a soft, dry cloth. Do not use abrasive cleaners or solvents, as these may damage the casing o componenti interni.

6.2 Conservazione

When not in use, store the SM-CAL1 in a cool, dry place, away from direct sunlight and extreme temperatures. If storing for extended periods, it is recommended to remove the 9V battery to prevent leakage.

6.3 Recalibration of the SM-CAL1

While the SM-CAL1 is designed for stable output, periodic verification of its accuracy by a certified calibration laboratory is recommended to ensure continued compliance with standards and optimal performance. The frequency of such recalibration depends on usage and regulatory requirements.

7. Risoluzione Dei Problemi

ProblemaPossibile causaSoluzione
No sound output / Power indicator offLa batteria è scarica o non è installata correttamente.Sostituire la batteria da 9 V, assicurandosi della corretta polarità.
Inaccurate calibration readings on sound meterMicrophone not properly seated; Calibrator needs recalibration; Sound meter issue.Ensure microphone is snugly fitted. Consider sending SM-CAL1 for professional recalibration. Check your sound meter's functionality.
Sound meter reading fluctuatesAmbient noise interference; Loose microphone connection.Perform calibration in a quiet environment. Re-seat the microphone firmly.

8. Specifiche

  • Modello: SM-CAL1
  • Output Sound Pressure Levels: 94 dB e 114 dB
  • Frequenza di uscita: 1,000 Hz (1 kHz)
  • Compatibilità microfono: Fits microphones up to 1/2 inch diameter
  • Precisione: Meets IEC 942 1988 Class 2 standards
  • Fonte di energia: One (1) 9V battery
  • Indicatore di batteria scarica:
  • Temperatura di esercizio: da 0°C a 40°C (da 32°F a 104°F)
  • Temperatura di conservazione: da -10°C a 50°C (da 14°F a 122°F)
  • Dimensions (H x D): Circa 4.9 x 1.9 pollici (125 x 48 mm)
  • Peso: Circa 0.6 libbre (285 g)
  • Certificazioni: CE compliant, ANSI, IEC standards

9. Garanzia e supporto

9.1 Informazioni sulla garanzia

Amprobe products are warranted against defects in material and workmanship for a specified period from the date of purchase. For detailed warranty terms and conditions, please refer to the warranty card included with your product or visit the official Ampvestaglia website. This warranty does not cover damage resulting from misuse, accident, neglect, alteration, contamination, or abnormal conditions of operation or handling.

9.2 Assistenza clienti

Se riscontri problemi o hai domande riguardanti il ​​tuo Amprobe SM-CAL1 Sound Meter Calibrator, please contact Amprobe customer support. Before contacting support, ensure you have reviewed the troubleshooting section of this manual. You can find contact information on the official Ampvestaglia websito o tramite il tuo distributore locale.

Documenti correlati - SM-CAL1

Preview AmpManuale utente del multimetro digitale robe AM-420
Manuale d'uso per il AmpMultimetro digitale Robe AM-420, che descrive in dettaglio le sue caratteristiche, le capacità di misurazione, le precauzioni di sicurezza e le specifiche per i test elettrici. Adatto per uso residenziale e professionale.
Preview AmpManuale utente del misuratore di capacità e resistenza Robe CR50A
Questo manuale utente fornisce istruzioni complete per il funzionamento del AmpMisuratore di capacità e resistenza Robe CR50A, che ne illustra le caratteristiche, le procedure di misurazione, le linee guida di sicurezza e le specifiche tecniche.
Preview AmpManuale utente del sistema di localizzazione di cavi/tubi sotterranei robe AT-3500
Manuale utente completo per il AmpSistema di localizzazione di cavi/tubi sotterranei robe AT-3500, che descrive in dettaglio funzionamento, caratteristiche, specifiche tecniche e manutenzione per un rilevamento accurato delle utenze sotterranee.
Preview AmpManuale di istruzioni del set di testa robe TT-200 - Test di linea e comunicazioni
Manuale di istruzioni completo per il Amprobe TT-200 Butt Set, che descrive in dettaglio le sue caratteristiche per test di linea, misurazione della resistenza di isolamento e comunicazioni temporanee per i tecnici.
Preview AmpManuale utente del termometro a infrarossi IR-750 / IR-750-EUR 50:1
Manuale d'uso per il AmpTermometri a infrarossi Robe IR-750 e IR-750-EUR 50:1. Le caratteristiche includono un ampio intervallo di temperatura, elevata precisione, puntatore laser, registrazione dei dati e connettività USB per applicazioni industriali e professionali.
Preview AmpGuida rapida al localizzatore di servizi sotterranei UAT-600 robe
Una guida rapida per l'avvio AmpSerie di localizzatori di servizi sotterranei UAT-600 robe, che fornisce istruzioni sulla localizzazione attiva e passiva dei servizi utilizzando varie modalità come alimentazione, radio, induzione, cavi di prova diretti e segnale Clamp.