Prodotto finitoview
The Hanna Instruments HI 99121N is a robust, portable, and waterproof pH and temperature meter designed for precise measurements in both diluted soil samples and direct soil applications. It comes equipped with the specialized HI 1292D pH electrode, making it an ideal tool for agricultural, environmental, and research professionals.

Image: Hanna Instruments HI 99121N pH/Temperature Meter with its HI 1292D electrode, showing the display and control buttons.
Key features include a large LCD screen for easy readability, automatic temperature compensation (ATC) for accurate readings across varying temperatures, a convenient hold function to freeze and record results, and a battery error protection system (BEPS) to ensure measurement reliability.
Caratteristiche principali:
- Portable, waterproof pH and temperature meter for soil and water applications.
- Automatic one- or two-point calibration for simplified setup and accuracy.
- Automatic Temperature Compensation (ATC) to eliminate manual calculations.
- Battery Error Protection System (BEPS) to prevent inaccurate readings due to low battery.
- Hold function to freeze the display and facilitate result recording.
Cosa c'è nella scatola
Upon unboxing your HI 99121N meter, please verify that all components listed below are present and in good condition:
- HI 99121N Meter
- HI 1292D pH/temperature electrode
- HI 721319 soil auger
- HI 7051M soil preparation solution
- HI 70004 pH 4.01 buffer solution sachet
- HI 70007 pH 7.01 buffer solution sachet
- HI 700663 cleaning solution sachet for inorganic soil deposits
- HI 700664 cleaning solution sachet for organic soil deposits
- 100 mL plastic beaker
- (3) batterie AAA da 1.5 V
- Manuale di istruzioni (questo documento)
- Custodia rigida per il trasporto
Impostazione e preparazione iniziale
- Installazione della batteria: Open the battery compartment cover on the back of the meter. Insert the three 1.5V AAA batteries, ensuring correct polarity. Close the cover securely.
- Collegamento dell'elettrodo: Connect the HI 1292D pH/temperature electrode to the BNC connector on the bottom of the meter. Ensure a firm and secure connection.
- Electrode Conditioning: Before first use, or after prolonged storage, condition the electrode by soaking its tip in HI 70300 storage solution or pH 7.01 buffer solution for at least 30 minutes. This ensures optimal performance.
- Calibrazione iniziale: It is recommended to perform an initial calibration before taking measurements. Refer to the "Operating Instructions - Calibration" section for detailed steps.
Istruzioni per l'uso
Accensione/spegnimento
- To turn the meter ON, premere il tasto MODALITÀ ON / OFF pulsante.
- To turn the meter SPENTO, premere e tenere premuto il tasto MODALITÀ ON / OFF button for approximately 2 seconds until the meter shuts down. The meter also features an auto-off function after 8 minutes of non-use to conserve battery life.
Calibrazione
The HI 99121N features automatic one- or two-point pH calibration. On-screen tutorial messages will guide you through the process.
- Ensure the electrode is clean and conditioned.
- Premere e tenere premuto il tasto MODALITÀ ON / OFF button for 5 seconds to enter calibration mode. The display will show "CAL" and "pH 7.01 USE".
- Rinse the electrode with distilled water and immerse it into the pH 7.01 buffer solution. Wait for the reading to stabilize. The meter will automatically recognize the buffer.
- If performing a two-point calibration, the display will then prompt for the second buffer (e.g., "pH 4.01 USE"). Rinse the electrode and immerse it into the pH 4.01 buffer solution.
- Once calibration is complete, the meter will return to measurement mode, displaying "Calibrated".
Taking Measurements (Soil and Water)
The HI 99121N can measure pH and temperature in both soil and water sampmeno.
For Direct Soil Measurement:
- For hard soil, use the HI 721319 soil auger to perforate the soil to the desired depth.
- Carefully insert the conical tip of the HI 1292D electrode directly into the soil. Ensure good contact between the electrode and the soil.
- Allow sufficient time for the reading to stabilize. The pH and temperature values will be displayed.
For Diluted Soil Samples or Water Samples:
- Prepara i tuoiample (e.g., diluted soil solution in the 100 mL beaker).
- Immergere la punta dell'elettrodo nel sample. Ensure the liquid level is above the sensing bulb.
- Gently stir the solution and wait for the reading to stabilize.
Funzione di attesa
Per congelare la lettura corrente sul display, premere il tasto SET HOLD button. The "HOLD" indicator will appear. Press the button again to release the hold and return to live measurement mode.
Manutenzione e cura
- Pulizia degli elettrodi: Regularly clean the HI 1292D electrode to ensure accurate readings. Use the provided HI 700663 (for inorganic deposits) or HI 700664 (for organic deposits) cleaning solution sachets as needed. Rinse thoroughly with distilled water after cleaning.
- Conservazione degli elettrodi: When not in use, always store the electrode with its protective cap filled with HI 70300 storage solution or pH 7.01 buffer solution. Never store the electrode dry or in distilled water.
- Sostituzione della batteria: When the battery indicator shows low power or the meter does not turn on, replace the three 1.5V AAA batteries. Ensure the meter is off before opening the battery compartment.
- Cura generale: Keep the meter and electrode clean and dry. Store the unit in its hard carrying case when not in use to protect it from dust and physical damage.
Risoluzione dei problemi
| Problema | Possibile causa | Soluzione |
|---|---|---|
| Inaccurate or Unstable Readings |
|
|
| Meter does not turn ON |
|
|
| "Err" or "Low Batt" message |
|
|
Specifiche
| Parametro | Valore |
|---|---|
| Intervallo di pH | -2.00 a 16.00 pH |
| Risoluzione del pH | PH 0.01 |
| Precisione del pH | ±0.02 pH |
| Intervallo di temperatura | -5.0 to 105.0 °C (23.0 to 221.0 °F) |
| Risoluzione della temperatura | 0.1 °C / 0.1 °F |
| Precisione della temperatura | ±0.5 °C (up to 60 °C), ±1 °C (outside); ±1.0 °F (up to 140 °F), ±2 °F (outside) |
| Compensazione della temperatura | Automatic (-5.0 to 105.0 °C / 23.0 to 221.0 °F) |
| Elettrodo | HI 1292D pH/Temperature Electrode (included) |
| Alimentazione elettrica | 3 batterie AAA da 1.5 V |
| Durata della batteria | Circa 1200 ore di utilizzo continuo |
| Auto-off | Dopo 8 minuti di non utilizzo |
| Dimensioni | 152 x 58 x 30 mm (6.0 x 2.3 x 1.2 pollici) |
| Peso | 205 g (7.2 once) |
Garanzia e supporto
Hanna Instruments products are manufactured to high standards and are warranted against defects in workmanship and materials. For specific warranty terms and conditions, please refer to the warranty card included with your product or visit the official Hanna Instruments websito.
For technical support, troubleshooting assistance, or to inquire about replacement parts and accessories, please contact Hanna Instruments customer service. Contact information can typically be found on the manufacturer's websito o nella confezione del prodotto.
Note: Unauthorized repairs or modifications may void the product warranty.





