Oakton WD-35811-71

Manuale di istruzioni per l'elettrodo pH Oakton WD-35811-71 Acorn All-in-One

For use with Oakton pH 6+ Meters

1. Introduzione

The Oakton WD-35811-71 Acorn All-in-One pH Electrode is a robust and reliable sensor designed for accurate pH measurements when paired with compatible Oakton pH 6+ meters. This electrode features a single junction, epoxy body, and is sealed for durability and minimal maintenance. This manual provides essential information for the proper setup, operation, and care of your pH electrode.

Oakton WD-35811-71 Acorn All-in-One pH Electrode

Image 1: Oakton WD-35811-71 Acorn All-in-One pH Electrode. This image displays the electrode, a cylindrical probe with a cable extending from one end, designed for pH measurement.

2. Informazioni sulla sicurezza

Always handle the pH electrode with care. The glass bulb at the tip is fragile. Avoid dropping or subjecting the electrode to mechanical shock. Do not allow the electrode to dry out, as this can permanently damage the sensor. Always store the electrode in its protective cap with appropriate storage solution.

3. Prodotto finitoview

The WD-35811-71 electrode is an all-in-one design, meaning it integrates both the pH sensing element and the reference electrode into a single probe. Its epoxy body provides enhanced durability, making it suitable for general laboratory and field use. The sealed design minimizes maintenance requirements.

Oakton WD-35811-71 pH Electrode Components

Immagine 2: Dettagliata view of the Oakton WD-35811-71 pH Electrode. This image shows the electrode from a different angle, highlighting the probe body, cable, and connector, providing a clearer view of its construction.

4. Impostazione

4.1 Disimballaggio

  • Carefully remove the electrode from its packaging.
  • Inspect the electrode for any visible damage. If damage is present, contact your supplier immediately.

4.2 Initial Conditioning

  • Remove the protective storage cap from the electrode tip. The cap typically contains a storage solution to keep the glass bulb hydrated.
  • Rinse the electrode tip with distilled or deionized water. Do not wipe the glass bulb, as this can create static charges and affect readings.
  • For optimal performance, condition the electrode by immersing the tip in pH 7 buffer solution for at least 30 minutes before first use.

4.3 Connecting to a pH Meter

  • Ensure your Oakton pH 6+ meter is turned off.
  • Carefully plug the electrode connector into the designated pH input port on your meter. Ensure a secure connection.
  • Accendere il pHmetro.

5. Operativo

This electrode is designed to work seamlessly with Oakton pH 6+ meters, utilizing their advanced features for accurate measurements.

5.1 Calibrazione

Regular calibration is crucial for accurate pH measurements. Refer to your Oakton pH 6+ meter's manual for specific calibration procedures. The WD-35811-71 electrode supports:

  • Three-point pH calibration: For enhanced accuracy across a wider pH range.
  • Auto buffer recognition: The meter automatically identifies standard pH buffer solutions.
  • mV offset calibration: Allows for fine-tuning of the electrode's millivolt response.

5.2 Esecuzione delle misurazioni

  • Sciacquare l'elettrodo con acqua distillata prima e dopo ogni misurazione.
  • Immergere la punta dell'elettrodo nel sample solution, ensuring the glass bulb and junction are fully submerged.
  • Stir the solution gently to ensure homogeneity and allow the reading to stabilize.
  • Il contatore Compensazione automatica della temperatura (ATC) feature will adjust readings for temperature variations, provided a temperature sensor is connected or temperature is manually entered.
  • Use the meter's Funzione di attesa per congelare la lettura corrente sul display.
  • Il contatore Auto-off feature will conserve battery life by automatically powering down after a period of inactivity.

6. Manutenzione

6.1 Pulizia dell'elettrodo

  • Dopo ogni utilizzo, sciacquare accuratamente l'elettrodo con acqua distillata o deionizzata.
  • If the electrode becomes fouled with oil, grease, or other contaminants, clean it with a mild detergent solution or a specialized electrode cleaning solution. Follow the cleaning solution manufacturer's instructions.
  • Non utilizzare mai materiali abrasivi o solventi forti per pulire l'elettrodo.

6.2 Conservazione

  • Conservare sempre l'elettrodo con il cappuccio protettivo riempito con la soluzione di conservazione dell'elettrodo pH (ad esempio, una soluzione di KCl 4M).
  • Never store the electrode dry or in distilled/deionized water, as this will dehydrate the glass bulb and shorten its lifespan.
  • Ensure the storage solution level covers the glass bulb and reference junction.

7. Risoluzione Dei Problemi

  • Unstable or Drifting Readings:
    • Ensure the electrode is properly conditioned and calibrated.
    • Check for air bubbles trapped around the glass bulb.
    • Clean the electrode if it appears fouled.
    • Verifica la sample temperature is stable.
  • Risposta lenta:
    • The electrode may be dehydrated; recondition in storage solution.
    • The electrode may be dirty; clean it thoroughly.
    • The electrode may be nearing the end of its lifespan.
  • No Reading or Error Message:
    • Check the connection between the electrode and the meter.
    • Ensure the meter is functioning correctly (test with another electrode if available).
    • Refer to your pH meter's manual for specific error codes.

8. Specifiche

ParametroSpecificazione
Numero di modelloWD-35811-71 (Internal: 1192M73EA)
TipoAll-in-One pH Electrode
GiunzioneSingle Junction
Materiale del corpoEpoxy
CompatibilitàOakton pH 6+ Meters
Features Supported (via meter)Automatic Temperature Compensation (ATC), Three-point pH calibration, Auto buffer recognition, mV offset calibration, Hold function, Auto-off
Dimensioni del pacco10.8 x 4 x 2.2 pollici
Peso0.02 once

9. Garanzia e supporto

For information regarding warranty coverage, technical support, or service, please refer to the documentation provided with your original purchase or contact Oakton Instruments directly through their official website or customer service channels. Please have your product model number (WD-35811-71) and purchase date available when contacting support.

Documenti correlati - WD-35811-71

Preview Istruzioni per l'uso del termometro a termocoppia serie Acorn OAKTON TEMP TC
Istruzioni per l'uso e specifiche per il termometro a termocoppia OAKTON TEMP TC Acorn Series, inclusi dettagli su misurazione, calibrazione, risoluzione dei problemi e accessori.
Preview Misuratori OAKTON ION Acorn Series: Istruzioni per l'uso e guida
Manuale operativo completo per i misuratori di pH/ioni OAKTON ION 5 e ION 6 della serie Acorn. Descrive la configurazione, la calibrazione, la misurazione, la manutenzione e la risoluzione dei problemi per un'analisi accurata della qualità dell'acqua.
Preview Timer OAKTON 410 con allarme: istruzioni per l'uso e specifiche
Istruzioni operative complete e specifiche per il timer con sveglia OAKTON 410 (modello WD-35002-00), che ne descrivono in dettaglio le caratteristiche, le funzioni dei pulsanti, le modalità di temporizzazione e le impostazioni della sveglia.
Preview Istruzioni per l'uso del cronometro OAKTON Single-Event 220 con allarme
Istruzioni operative dettagliate per il cronometro OAKTON Single-Event 220 con sveglia (modello WD-35002-10), che coprono le funzioni del cronometro, le impostazioni di ora/calendario, le funzioni di sveglia, le specifiche e le informazioni sulla garanzia.
Preview Manuale di istruzioni del misuratore di ossigeno disciolto Oakton DO250/DO260
Manuale di istruzioni completo per i misuratori di ossigeno disciolto Oakton DO250 e DO260, che copre il prodotto oltreview, operazioni di base, calibrazione, gestione dei dati, configurazione, manutenzione, risoluzione dei problemi e precauzioni di sicurezza.
Preview Manuale di istruzioni per i misuratori portatili della qualità dell'acqua della serie OAKTON 300
Questo manuale di istruzioni fornisce una guida completa per il funzionamento dei misuratori portatili della qualità dell'acqua della serie OAKTON 300. Copre il prodotto oltreview, setup, calibration, measurement, data management, maintenance, and troubleshooting for accurate water quality analysis.