Introduzione
This manual provides comprehensive instructions for the setup, operation, maintenance, and troubleshooting of the Oakton Instruments WD-35419-10 Series pH 700 Benchtop Meter. It is designed to assist users in achieving accurate and reliable pH measurements in laboratory and industrial settings. Please read this manual thoroughly before operating the instrument.
Contenuto della confezione
Verificare che tutti gli articoli siano presenti e integri al momento del disimballaggio:
- pH 700 Benchtop Meter unit
- Double-Junction Glass pH Electrode
- Automatic Temperature Compensation (ATC) Probe
- Electrode Stand
- Power Adapter (110/220 VAC compatible)
- Manuale di istruzioni (questo documento)
Prodotto finitoview

Figure 1: Oakton Instruments WD-35419-10 Series pH 700 Benchtop Meter. The image displays the meter's digital display unit, a double-junction glass pH electrode immersed in a green liquid in a beaker, and an automatic temperature compensation (ATC) probe, all supported by an adjustable stand.
The Oakton Instruments pH 700 Benchtop Meter is designed for precise pH and temperature measurements. Its compact footprint makes it suitable for various laboratory environments. Key features include a large LCD display, intuitive button interface, and compatibility with a double-junction pH electrode for enhanced accuracy and longevity.
Impostare
- Disimballare i componenti: Rimuovere con attenzione tutti i componenti dalla confezione.
- Posizionare il supporto: Place the electrode stand on a stable, level surface next to the meter.
- Collegare l'elettrodo: Connect the double-junction glass pH electrode to the BNC connector on the back of the meter. Ensure a secure connection.
- Connect the ATC Probe: Plug the ATC probe into its designated port on the meter.
- Connetti alimentazione: Plug the power adapter into the meter's power input, then connect it to a 110/220 VAC power outlet.
- Accensione iniziale: Press the power button to turn on the meter. The display should illuminate.
Istruzioni per l'uso
Calibrazione
Regular calibration is essential for accurate pH measurements. Calibrate the meter using at least two standard pH buffer solutions (e.g., pH 4.01, 7.00, and 10.00).
- Preparare i buffer: Ensure buffer solutions are at room temperature.
- Elettrodo di risciacquo: Rinse the pH electrode and ATC probe with distilled water and gently blot dry.
- Inizia la calibrazione: Refer to the meter's on-screen prompts or specific button sequence (usually a "CAL" button) to enter calibration mode.
- First Buffer: Immerse the electrode and ATC probe into the first buffer solution (e.g., pH 7.00). Wait for the reading to stabilize, then confirm.
- Second Buffer: Rinse the electrode and probe, then immerse them into the second buffer solution (e.g., pH 4.01 or 10.00). Wait for stabilization and confirm.
- Calibrazione completa: The meter will indicate when calibration is complete.
Prendere Misure
- Elettrodo di risciacquo: Rinse the calibrated electrode and ATC probe with distilled water.
- Immerse in Sampon: Immerse the electrode and ATC probe into the sample solution. Ensure the electrode junction and temperature sensor are fully submerged.
- Stabilizzare la lettura: Allow sufficient time for the reading to stabilize. The meter will typically display a "READY" indicator when stable.
- Lettura del record: Note the pH and temperature values displayed.
- Pulire dopo l'uso: Rinse the electrode and probe thoroughly with distilled water after each measurement.
Manutenzione
- Conservazione degli elettrodi: Always store the pH electrode in its protective cap filled with electrode storage solution or pH 7 buffer. Never store in distilled water.
- Pulizia: Clean the electrode regularly to prevent buildup. For general cleaning, soak in a mild detergent solution, then rinse thoroughly. For stubborn deposits, consult the electrode's specific cleaning instructions.
- Pulizia del contatore: Pulisci l'esterno del contatore con un pannoamp stoffa. Non utilizzare prodotti chimici aggressivi o detergenti abrasivi.
- Frequenza di calibrazione: Calibrate the meter daily or before each use for optimal accuracy, especially when measuring samples with significantly different pH values.
Risoluzione dei problemi
| Problema | Possibile causa | Soluzione |
|---|---|---|
| Letture instabili | Dirty electrode; air bubbles; electrode not fully immersed; expired buffer; damaged electrode. | Clean electrode; gently tap electrode to remove bubbles; ensure full immersion; use fresh buffer; replace electrode. |
| Letture imprecise | Improper calibration; expired buffer; electrode not conditioned; temperature difference. | Recalibrate with fresh buffers; condition electrode; ensure ATC probe is functioning. |
| Il misuratore non si accende | Power adapter disconnected; faulty power outlet; meter malfunction. | Check power connections; try a different outlet; contact technical support. |
Specifiche
- Numero modello: WD-35419-10
- Dimensioni del prodotto: 12.3 x 9.2 x 5 pollici
- Peso: 4 libbre
- Requisiti di alimentazione: 110 / 220 VAC
- Tipo di elettrodo: Double-Junction Glass pH Electrode
- Compensazione della temperatura: Automatic (with ATC Probe)
- Produttore: Thermo Scientific (as listed on product page)
- Paese di origine: Cina
- Orma: Compact, over 40% smaller than comparable benchtop meters.
Garanzia e supporto
For warranty information, technical support, or service inquiries, please refer to the documentation provided with your purchase or contact Oakton Instruments customer service directly. Keep your purchase receipt as proof of purchase.