WTW 202706

WTW 202706 Heavy-Duty DO Probe Maintenance Kit User Manual

For CellOx Probes

1. Introduzione

This manual provides comprehensive instructions for the proper use and maintenance of the WTW 202706 Heavy-Duty DO Probe Maintenance Kit. This kit is designed to ensure the optimal performance and longevity of your WTW CellOx Dissolved Oxygen (DO) probes.

The kit includes essential components for routine care, such as cleaning solution, abrasive foil, membrane heads, and electrolyte solution, all crucial for accurate and reliable dissolved oxygen measurements.

2. Contenuto del kit

The WTW 202706 Heavy-Duty DO Probe Maintenance Kit contains the following items:

  • Cleaning solution for oxygen sensors (30 ml)
  • Electrolyte solution for oxygen sensors (50 ml)
  • Abrasive foil
  • Three membrane heads
Contents of the WTW 202706 Heavy-Duty DO Probe Maintenance Kit, including two bottles of solution, a clear protective cap, a black probe component, and a yellow protective tube.

Figura 1: Oltreview of the WTW 202706 Heavy-Duty DO Probe Maintenance Kit components. Visible are bottles of cleaning and electrolyte solutions, a protective cap, a probe component, and a yellow protective tube.

3. Impostazione e preparazione

Before performing maintenance, ensure you have a clean, dry workspace and any necessary personal protective equipment (PPE).

  1. Identify Probe Type: Confirm that your DO probe is a WTW CellOx probe, as this kit is specifically designed for these models.
  2. Raccogli i componenti: Lay out all components from the maintenance kit and your DO probe.
  3. Leggi le istruzioni: Familiarize yourself with the entire maintenance procedure before beginning.

4. Operating Procedures (Probe Maintenance)

Regular maintenance is crucial for accurate DO measurements. Follow these steps for proper probe care:

4.1. Pulizia della sonda

  1. Sonda di risciacquo: Gently rinse the probe with distilled water to remove any loose debris.
  2. Applicare la soluzione detergente: Apply a small amount of the provided cleaning solution to a soft cloth or directly to the probe's membrane area.
  3. Pulisci delicatamente: Carefully wipe the membrane and probe body. For stubborn deposits, use the abrasive foil provided, but apply minimal pressure to avoid damaging the membrane.
  4. Risciacquare abbondantemente: Rinse the probe thoroughly with distilled water after cleaning to remove all traces of the cleaning solution.

4.2. Replacing the Membrane Head

The membrane head should be replaced periodically or if damaged.

  1. Svitare la vecchia testa: Carefully unscrew the old membrane head from the probe body.
  2. Prepare New Head: Take one of the new membrane heads from the kit.
  3. Fill with Electrolyte: Fill the new membrane head with the provided electrolyte solution. Ensure there are no air bubbles trapped inside. Gently tap the head to dislodge any bubbles.
  4. Screw On New Head: Carefully screw the new membrane head onto the probe body. Ensure it is finger-tight to prevent leaks, but do not overtighten.

4.3. Electrolyte Replacement

The electrolyte solution should be replaced when replacing the membrane head or if the probe's response becomes sluggish.

  1. Drain Old Electrolyte: If not replacing the membrane head, carefully drain the old electrolyte solution from the existing membrane head.
  2. Refill with Fresh Electrolyte: Fill the membrane head with fresh electrolyte solution from the kit, ensuring no air bubbles are present.

5. Stoccaggio e movimentazione

Proper storage of the kit components and the probe itself ensures their longevity and readiness for use.

  • Store the kit components in a cool, dry place, away from direct sunlight and extreme temperatures.
  • Keep bottles of cleaning and electrolyte solutions tightly sealed when not in use.
  • Always store the DO probe with its protective cap filled with a small amount of electrolyte solution or in a suitable storage solution as recommended by WTW.

6. Risoluzione dei problemi comuni

If you encounter issues with your DO probe after maintenance, consider the following:

ProblemaPossibile causaSoluzione
Letture instabiliAir bubbles in membrane head, dirty membrane, depleted electrolyte.Tap membrane head to remove bubbles, clean membrane, replace electrolyte.
Tempo di risposta lentoOld or contaminated electrolyte, damaged membrane.Replace electrolyte and/or membrane head.
Nessuna letturaCompletely dry membrane, severe probe damage.Refill with electrolyte and re-condition probe. If still no reading, probe may need replacement.

7. Specifiche

Key specifications for the WTW 202706 Heavy-Duty DO Probe Maintenance Kit:

  • Numero modello: 202706
  • Compatibilità: Designed for WTW CellOx DO probes
  • Contenuto del kit: Cleaning solution, abrasive foil, 3 membrane heads, electrolyte solution
  • Dimensioni della confezione: 8.79 x 5.49 x 2.64 pollici
  • Peso del pacco: 9.44 once
  • Produttore: COLE-PARMER (for distribution)
  • Marca: Che succede?

8. Garanzia e supporto

The WTW 202706 Heavy-Duty DO Probe Maintenance Kit comes with a Garanzia di 1 anni dalla data di acquisto.

For technical support, warranty claims, or further assistance, please contact your authorized WTW distributor or visit the official WTW website. Ensure you have your product model number (202706) and proof of purchase available when seeking support.

Documenti correlati - 202706

Preview WTW-PDRP4620EASD3/4630ESD3 カメラ 取扱説明書:メニュー設定と安全ガイド
この取扱説明書は、WTW-PDRP4620EASD3 および WTW-PDRP4630ESD3 ドームカメラのOSDメニュー設定、基本接続、および安全上の注意事項を解説します。製品のセットアップと安全な使用に必要な情報を提供します。
Preview WTW-1213NXシリーズ IPカメラ 取扱説明書 - WTW-EAGLE-VIPアプリ設定ガイド
塚本無線 WTW-1213NXシリーズ IPカメラの取扱説明書。WTW-EAGLE-VIPアプリのインストール、カメラの設置、設定、録画、共有方法などを解説。
Preview WTW Camera OSD Menu and Connection Guide
Comprehensive guide to connecting WTW surveillance cameras and configuring their On-Screen Display (OSD) menu settings, covering video modes, exposure, day/night settings, and more. Includes detailed connection diagrams and setup instructions.
Preview WTW セキュリティカメラシステム 取扱説明書:安全上の注意とメニュー設定
WTW製セキュリティカメラシステムの安全上の注意、基本接続方法、およびOSDメニュー設定に関する詳細な取扱説明書。カメラ設定、ビデオモード、言語選択、露出調整などの機能について解説。
Preview WTW-WSP12-BM / WSP18-BM / WSP21-BM IPカメラ 取扱説明書
塚本無線 WTW-WSP12-BM, WTW-WSP18-BM, WTW-WSP21-BM IPカメラの取扱説明書。設置、設定、機能、トラブルシューティング、保証に関する詳細情報。
Preview WTW-EGDRY388GWX/GBX IP Camera User Manual
This manual provides instructions for setting up, registering, and using the WTW-EGDRY388GWX/GBX IP Camera, including app installation, camera configuration, motion detection, recording, playback, sharing, and troubleshooting.