Oakton AO-35613-22

Manuale utente del misuratore portatile Oakton AO-35613-22 pH 6+

Modello: AO-35613-22

1. Introduzione

This manual provides instructions for the proper use and maintenance of the Oakton AO-35613-22 pH 6+ Handheld Meter. This portable device is designed for accurate pH measurements in various applications. Please read this manual thoroughly before operating the meter.

2. Informazioni sulla sicurezza

3. Contenuto della confezione

Verificare che tutti gli articoli siano presenti al momento del disimballaggio:

4. Prodotto finitoview

The Oakton AO-35613-22 is a portable pH meter designed for accurate and rapid measurements. It features an LCD display, intuitive controls, and automatic temperature compensation.

Oakton AO-35613-22 pH 6+ Handheld Meter with pH probe and temperature probe connected

Figura 1: Oakton AO-35613-22 pH 6+ Handheld Meter. The image shows the meter with its LCD display showing '7.00 pH', and connected pH and temperature probes. The meter is encased in a protective rubber boot.

Caratteristiche principali:

5. Impostazione

5.1 Installazione della batteria

  1. Rimuovere la cuffia protettiva in gomma.
  2. Individuare il coperchio del vano batterie sul retro del misuratore.
  3. Open the cover and insert four (4) AAA batteries, observing polarity.
  4. Replace the battery compartment cover and the protective rubber boot.

5.2 Collegamento della sonda

  1. Connect the pH probe to the BNC connector on the top of the meter.
  2. Connect the ATC probe to the mini-jack connector next to the BNC connector.
  3. Assicurati che le connessioni siano sicure.

6. Istruzioni per l'uso

6.1 Accensione/spegnimento

6.2 Esecuzione di una misurazione

  1. Ensure the meter is calibrated (refer to Section 7).
  2. Rinse the pH probe and ATC probe with distilled water.
  3. Immerse both probes into the sample solution. Ensure the probe tips are fully submerged.
  4. Allow the reading to stabilize on the LCD display.
  5. Premere il tasto HOLD/ENTER button to freeze the reading if desired. Press again to release.
  6. After measurement, rinse probes thoroughly with distilled water and store them properly.

6.3 Selezione della modalità

Premere il tasto MODE/°C/°F button to switch between pH, mV, and temperature display modes.

7. Calibrazione

Regular calibration is essential for accurate pH measurements. The meter supports up to 5-point calibration with NIST and USA buffer recognition.

  1. Prepare fresh pH buffer solutions (e.g., pH 4.01, 7.00, 10.01).
  2. Rinse the pH and ATC probes with distilled water.
  3. Immerse both probes into the first buffer solution (e.g., pH 7.00).
  4. Premere il tasto CAL button. The meter will automatically recognize the buffer and calibrate.
  5. Once stable, rinse probes and repeat the process for subsequent buffer solutions (e.g., pH 4.01, then pH 10.01).
  6. Refer to the full calibration procedure in the detailed manual for advanced settings and multi-point calibration.

8. Manutenzione e cura

8.1 Pulizia

8.2 Conservazione

9. Risoluzione Dei Problemi

ProblemaPossibile causaSoluzione
Nessuna visualizzazioneBatterie scariche o installate in modo errato.Controllare la polarità delle batterie; sostituire le batterie.
Letture instabiliProbe not immersed correctly; dirty probe; old buffer solutions; damaged probe.Ensure proper immersion; clean probe; use fresh buffer solutions; replace probe if damaged.
Errore di calibrazioneIncorrect buffer solution; dirty probe; expired buffer.Verify buffer values; clean probe; use fresh buffer solutions.

10. Specifiche

Tipo di prodottoMisuratore di pH
Meter StylePortatile
Tipo di visualizzazioneLCD
Intervallo di pHda 0.00 a 14.00 pH
Risoluzione del pH0.01 o 0.1 pH
Precisione del pH±0.01 pH
Gamma mV±1999 mV
mV Resolution1 mV
mV Accuracy±0.2 mV or 0.05% (whichever is greater)
Intervallo di temperatura-10 a 110 °C
Compensazione della temperaturaAutomatic (0 to 100 °C)
Buffer RecognitionNIST and USA
Punti di calibrazioneFino a 5 punti
Fonte di alimentazione4 batterie AAA
Durata della batteriaCirca 700 ore
Dimensioni (L x A)8.57 cm x 4.29 cm (3 3/8 in x 1 11/16 in)
Materiale esternoRubber (protective boot)

11. Informazioni sulla garanzia

The Oakton AO-35613-22 pH 6+ Handheld Meter comes with a Garanzia di 3 anno from the date of purchase. This warranty covers defects in materials and workmanship under normal use. Please retain your proof of purchase for warranty claims. The warranty does not cover damage caused by misuse, accident, unauthorized modification, or improper storage of the pH probe.

12. Supporto e contatto

For technical assistance, troubleshooting beyond this manual, or warranty service, please contact Oakton customer support or visit the official Oakton websito. Le informazioni di contatto si trovano in genere sulla confezione del prodotto o sul sito del produttore websito.

Manufacturer: Cole-Parmer

Websito: Visit the Oakton Store on Amazon

Documenti correlati - AO-35613-22

Preview Misuratori OAKTON ION Acorn Series: Istruzioni per l'uso e guida
Manuale operativo completo per i misuratori di pH/ioni OAKTON ION 5 e ION 6 della serie Acorn. Descrive la configurazione, la calibrazione, la misurazione, la manutenzione e la risoluzione dei problemi per un'analisi accurata della qualità dell'acqua.
Preview Catalogo di misuratori e tester per la qualità dell'acqua Oakton | Cole-Parmer
Esplora la gamma completa di misuratori e tester della qualità dell'acqua Oakton di Cole-Parmer, dotati di soluzioni avanzate per la misurazione di pH, conduttività e ORP per analisi accurate.
Preview Manuale di istruzioni del misuratore di ossigeno disciolto Oakton DO250/DO260
Manuale di istruzioni completo per i misuratori di ossigeno disciolto Oakton DO250 e DO260, che copre il prodotto oltreview, operazioni di base, calibrazione, gestione dei dati, configurazione, manutenzione, risoluzione dei problemi e precauzioni di sicurezza.
Preview Manuale di istruzioni per il misuratore portatile della qualità dell'acqua Oakton serie 300
Manuale di istruzioni per i misuratori portatili della qualità dell'acqua Oakton serie 300 (modelli 350, 360, 380). Descrive funzionamento, calibrazione, manutenzione, sicurezza e specifiche per un'analisi accurata dell'acqua.
Preview Manuale di istruzioni per i misuratori portatili della qualità dell'acqua della serie OAKTON 300
Questo manuale di istruzioni fornisce una guida completa per il funzionamento dei misuratori portatili della qualità dell'acqua della serie OAKTON 300. Copre il prodotto oltreview, setup, calibration, measurement, data management, maintenance, and troubleshooting for accurate water quality analysis.
Preview Istruzioni per l'uso del misuratore di pH/ORP/TEMP OAKTON pH 700
Questo documento fornisce istruzioni operative complete per il misuratore di pH/ORP/TEMP OAKTON pH 700, che ne descrivono le caratteristiche, la calibrazione, le procedure di misurazione, la risoluzione dei problemi e le specifiche tecniche.