Oakton WD-35805-04

Manuale di istruzioni per l'elettrodo pH da laboratorio Oakton WD-35805-04

Modello: WD-35805-04

Introduzione

This manual provides comprehensive instructions for the proper setup, operation, and maintenance of your Oakton WD-35805-04 Laboratory pH Electrode. Adhering to these guidelines will ensure optimal performance and longevity of the device.

Prodotto finitoview

The Oakton WD-35805-04 is a double-junction, glass body, refillable pH electrode designed for laboratory applications. Its robust construction and double-junction design make it suitable for a wide range of chemical measurements, offering enhanced durability and resistance to aggressive chemicals. This electrode features an Ag/AgCl reference cell for stable and accurate readings.

Oakton WD-35805-04 Laboratory pH Electrode

Image 1: Oakton WD-35805-04 Laboratory pH Electrode. This image shows the full electrode, featuring a black top housing with an "OPEN" indicator, a clear glass body revealing internal components, and a blue spherical tip. A black cable extends from the top.

Caratteristiche principali:

Impostare

  1. Disimballaggio: Estrarre con cautela l'elettrodo dalla confezione. Verificare la presenza di eventuali danni visibili.
  2. Electrolyte Filling (if necessary): This electrode is refillable. If the electrolyte level is low, open the fill hole (usually indicated by an "OPEN" arrow or cap) and carefully add the appropriate reference electrolyte solution (e.g., 3M KCl) until the level is just below the fill hole. Close the fill hole securely.
  3. Risciacquo: Rinse the electrode tip with distilled or deionized water to remove any storage solution. Do not wipe the glass bulb, as this can create static charges.
  4. Condizionata: For optimal performance, soak the electrode in a pH 7 buffer solution for at least 30 minutes (or longer, up to several hours) before first use or after prolonged dry storage. This rehydrates the glass membrane.
  5. Connecting to Meter: Connect the electrode's BNC connector (or other specified connector type) to the corresponding input on your pH meter. Ensure a secure connection.
  6. Calibrazione: Calibrate your pH meter with the electrode using at least two, preferably three, fresh pH buffer solutions (e.g., pH 4.01, 7.00, and 10.00). Follow your pH meter's specific calibration procedure.

Istruzioni per l'uso

  1. Preparazione: Ensure the electrode is properly conditioned and calibrated.
  2. Sample Misurazione:
    • Risciacquare l'elettrodo con acqua distillata o deionizzata tra un'operazione e l'altra.ampmeno.
    • Immergere la punta dell'elettrodo nel sample solution, ensuring the glass bulb and junction are fully submerged.
    • Stir the solution gently to ensure homogeneity and faster response.
    • Allow the reading to stabilize before recording the pH value. Stability is typically indicated by a stable reading on the pH meter for a few seconds.
  3. Compensazione della temperatura: If your pH meter has automatic temperature compensation (ATC), ensure the temperature probe is also immersed in the sample. If not, manually input the sample temperature into the meter.
  4. Post-Measurement: Dopo ogni misurazione, sciacquare accuratamente l'elettrodo con acqua distillata o deionizzata.

Manutenzione

Risoluzione dei problemi

ProblemaPossibile causaSoluzione
Risposta lenta / Letture instabili
  • Membrana di vetro disidratata
  • Giunzione intasata
  • Basso livello di elettroliti
  • Soluzioni tampone contaminate
  • Temperatura non stabile
  • Recondition electrode in pH 7 buffer.
  • Clean electrode as per Maintenance section.
  • Refill electrolyte.
  • Use fresh buffer solutions.
  • Consenteample/electrode to reach thermal equilibrium.
Letture imprecise
  • Calibrazione impropria
  • Elettrodo sporco
  • Expired or contaminated buffers
  • Electrode nearing end of life
  • Recalibrate with fresh buffers.
  • Clean electrode.
  • Replace buffers.
  • Consider replacing electrode if issues persist after cleaning/reconditioning.
No Reading / "Error" on Meter
  • Collegamento allentato
  • Elettrodo rotto
  • Malfunzionamento del misuratore
  • Check electrode connection to meter.
  • Inspect electrode for visible damage (cracks in glass).
  • Test meter with a known good electrode or consult meter manual.

Specifiche

AttributoDettaglio
Numero di modelloWD-35805-04
Tipo di elettrodoLaboratory pH Electrode, Double-Junction
Materiale del corpoBicchiere
Tipo di riferimentoRefillable Ag/AgCl
Dimensioni del pacco10.9 x 4.1 x 2.2 pollici
Peso dell'articolo6.38 once
ProduttoreStrumenti OAKTON
Data prima disponibilità2 giugno 2016
ASINB01GIRDK64
Classifica dei best-sellerSee Top 100 in Industrial & Scientific

Garanzia e supporto

For specific warranty information regarding your Oakton WD-35805-04 Laboratory pH Electrode, please refer to the documentation included with your purchase or visit the official Oakton Instruments website. For technical support, troubleshooting assistance beyond this manual, or to inquire about replacement parts and accessories, please contact Oakton Instruments customer service directly.

Oakton Instruments Contact Information: (Fare riferimento alla confezione del prodotto o alle informazioni ufficiali) websito per i dettagli di contatto aggiornati.)

Documenti correlati - WD-35805-04

Preview Timer OAKTON 410 con allarme: istruzioni per l'uso e specifiche
Istruzioni operative complete e specifiche per il timer con sveglia OAKTON 410 (modello WD-35002-00), che ne descrivono in dettaglio le caratteristiche, le funzioni dei pulsanti, le modalità di temporizzazione e le impostazioni della sveglia.
Preview Istruzioni per l'uso del termometro a termocoppia serie Acorn OAKTON TEMP TC
Istruzioni per l'uso e specifiche per il termometro a termocoppia OAKTON TEMP TC Acorn Series, inclusi dettagli su misurazione, calibrazione, risoluzione dei problemi e accessori.
Preview Misuratori OAKTON ION Acorn Series: Istruzioni per l'uso e guida
Manuale operativo completo per i misuratori di pH/ioni OAKTON ION 5 e ION 6 della serie Acorn. Descrive la configurazione, la calibrazione, la misurazione, la manutenzione e la risoluzione dei problemi per un'analisi accurata della qualità dell'acqua.
Preview Istruzioni per l'uso del cronometro OAKTON Single-Event 220 con allarme
Istruzioni operative dettagliate per il cronometro OAKTON Single-Event 220 con sveglia (modello WD-35002-10), che coprono le funzioni del cronometro, le impostazioni di ora/calendario, le funzioni di sveglia, le specifiche e le informazioni sulla garanzia.
Preview Manuale di istruzioni del misuratore di ossigeno disciolto Oakton DO250/DO260
Manuale di istruzioni completo per i misuratori di ossigeno disciolto Oakton DO250 e DO260, che copre il prodotto oltreview, operazioni di base, calibrazione, gestione dei dati, configurazione, manutenzione, risoluzione dei problemi e precauzioni di sicurezza.
Preview Manuale di istruzioni per i misuratori portatili della qualità dell'acqua della serie OAKTON 300
Questo manuale di istruzioni fornisce una guida completa per il funzionamento dei misuratori portatili della qualità dell'acqua della serie OAKTON 300. Copre il prodotto oltreview, setup, calibration, measurement, data management, maintenance, and troubleshooting for accurate water quality analysis.