Introduzione
The Traceable Excursion-Trac Data Logger Thermometer is designed for precise temperature monitoring and data recording. This device simultaneously displays current, minimum, and maximum temperatures, and features a rolling memory for extensive data storage. It is equipped with a submersible probe and offers customizable alarm functions to ensure critical temperature conditions are maintained and recorded.

Image: The Traceable Excursion-Trac Data Logger Thermometer, showing its digital display and connected bullet probe.
Contenuto della confezione
Verificare che tutti gli articoli siano presenti al momento del disimballaggio:
- Traceable Excursion-Trac Data Logger Thermometer Unit
- Bullet Probe with 10-ft (3-m) Cable
- Chiavetta USB
- Flip-Open Stand
- Hook-and-Loop Strips
- Magnetic Strips
- Wall-Mount Receptacle
- Batterie (preinstallate o separate)
- NIST-Traceable Calibration Certificate with Data
Impostare
- Installazione della batteria: If batteries are not pre-installed, open the battery compartment on the back of the unit and insert the specified batteries, observing correct polarity.
- Connessione sonda: Connect the bullet probe cable to the designated port on the thermometer unit. Ensure a secure connection.
- Posizionamento iniziale: Position the thermometer unit using the flip-open stand for a benchtop, or utilize the hook-and-loop strips, magnetic strips, or wall-mount receptacle for secure placement. The 10-ft (3-m) submersible micro-cable allows the probe to be placed in the desired monitoring location, even permitting doors to close over the cable.
- Accensione: Press the power button (if available) or ensure the unit is active. The display should illuminate, showing current temperature readings.
Istruzioni per l'uso
Funzionamento di base
- Display: The LCD simultaneously shows the current temperature, minimum recorded temperature (MIN), and maximum recorded temperature (MAX).
- Unità: Use the 'UNITS' button to toggle between Celsius (°C) and Fahrenheit (°F) temperature scales.
- Memoria View: The 'MEM VIEW' function allows review dei dati memorizzati.
Registrazione dei dati
- The device features rolling memory, capable of retaining the most recent 525,600 temperature observations.
- Trasferimento dati: To transfer recorded data, insert the provided USB flash drive into the unit's USB port. The data will be saved as a CSV file, accessible on any computer without special software.
- Cancella memoria: After downloading data and alarm events, the memory can be cleared to continue recording new data. Refer to the device's specific buttons (e.g., 'CLEAR') for this function.
Funzioni di allarme
- The thermometer allows setting limits for and storing up to 10 unique alarm events.
- Impostazione degli allarmi: Use the 'ALARM' and 'SELECT/ADVANCE' buttons to set desired high and low temperature alarm thresholds.
- Smart-Alarm Function: The Smart-Alarm provides both visual and audible alerts. It continues to alert even if conditions return to the normal temperature range, ensuring that any excursion is noted.
- Silenzia allarme: An 'ALARM SILENCE' button is available to temporarily mute audible alerts.
Manutenzione
- Pulizia: Pulisci l'unità con l'annuncioamp cloth. Do not submerge the main unit in water. The bullet probe is submersible and can be cleaned as needed.
- Sostituzione della batteria: When the battery indicator appears or the unit fails to power on, replace the batteries. Refer to the battery compartment for the correct battery type.
- Cura della sonda: Handle the probe and its cable with care to prevent damage. Avoid sharp bends or excessive pulling on the cable.
Risoluzione dei problemi
| Problema | Possibile causa | Soluzione |
|---|---|---|
| L'unità non si accende o il display è vuoto. | Batterie scariche o installate in modo non corretto. | Sostituire le batterie assicurandosi che la polarità sia corretta. |
| Letture della temperatura imprecise. | Sonda non collegata correttamente o danneggiata. | Ensure probe is securely connected. Inspect probe and cable for visible damage. |
| Data not transferring to USB drive. | USB drive not properly inserted or full. | Re-insert USB drive. Ensure the drive has sufficient space. Try a different USB drive if available. |
| Alarm not sounding or not clearing. | Alarm settings incorrect or alarm silenced. | Verify alarm thresholds. Check if the alarm has been manually silenced. |
Specifiche
| Caratteristica | Dettaglio |
|---|---|
| Numero di modello | Excursion-Trac |
| Intervallo di temperatura | –58 to 158 Degrees Fahrenheit (-50 to 70 Degrees Celsius) |
| Risoluzione | 0.01 gradi |
| Precisione | 0.25 grado Celsius |
| Capacità di registrazione dei dati | 525,600 temperature observations (rolling memory) |
| Connettività | USB (for CSV data transfer) |
| Tipo di visualizzazione | LCD |
| Caratteristica speciale | Alarm (Smart-Alarm function) |
| Tipo di sonda | Bullet Probe, Submersible, 10-ft (3-m) cable |
| Materiale | Plastica |
| Cura del prodotto | Pulisci Pulire |
| Certificazione | Taratura tracciabile NIST |
Supporto e garanzia
For technical assistance, product inquiries, or warranty information, please contact the manufacturer, Cole-Parmer, or visit their official website. Specific warranty details were not provided in the product data.
Produttore: COLE-PARMER
Marca: MATERIA
Supporto online: www.coleparmer.com