OFITE API Filter Press with CO2 Pressure Assembly

Manuale utente del filtro pressa API con gruppo di pressione CO2

Model: API Filter Press with CO2 Pressure Assembly

Marca: OFITE

1. Introduzione

The OFITE API Filter Press with CO2 Pressure Assembly is a precision instrument designed for evaluating the filtration properties of drilling fluids. This device provides a quick and easy method to determine fluid loss and filter cake characteristics under controlled pressure conditions. It is engineered to conform to American Petroleum Institute (API) specifications, making it suitable for both field and laboratory applications in the oil and gas industry.

Understanding the filtration behavior of drilling fluids is crucial for optimizing drilling operations, preventing formation damage, and ensuring wellbore stability. This manual provides comprehensive instructions for the safe and effective use, maintenance, and troubleshooting of your API Filter Press.

2. Informazioni sulla sicurezza

Per evitare lesioni e danni all'attrezzatura, attenersi sempre alle seguenti linee guida di sicurezza:

3. Componenti finitiview

OFITE API Filter Press with CO2 Pressure Assembly
Figura 3.1: OFITE API Filter Press with CO2 Pressure Assembly. This image shows the complete filter press unit, including the pressure cell, frame, CO2 pressure regulator with gauge, and the graduated cylinder for collecting filtrate. A box of filter paper is visible at the base.

4. Istruzioni per l'installazione

  1. Disimballaggio e ispezione: Carefully unpack all components. Inspect for any shipping damage. Report any damage to the supplier immediately.
  2. Montaggio:
    • Place the filter press frame on a stable, level surface.
    • Attach the CO2 pressure regulator assembly to the pressure cell inlet, ensuring a tight seal.
    • Connect the CO2 pressure regulator to a suitable CO2 cylinder. Ensure the cylinder is secured to prevent tipping.
  3. Preparation of Pressure Cell:
    • Ensure the pressure cell is clean and dry.
    • Place a new filter paper (e.g., OFITE Catalog No. 140-55) at the bottom of the pressure cell, ensuring it lies flat and covers the screen.
    • Secure the bottom cap of the pressure cell, ensuring the O-ring is properly seated for a leak-proof seal.
  4. Fluido Sampla Preparazione: Preparare il fluido di perforazioneample according to relevant API recommended practices or your specific testing protocol.

5. Istruzioni per l'uso

Follow these steps for conducting a filtration test:

  1. Carichiampon: Pour the prepared drilling fluid sample into the pressure cell. Do not overfill.
  2. Seal Cell: Securely attach the top cap of the pressure cell, ensuring the O-ring is properly seated.
  3. Position Graduated Cylinder: Place a clean, dry graduated cylinder beneath the filtrate outlet of the pressure cell.
  4. Applicare pressione:
    • Slowly open the valve on the CO2 cylinder.
    • Adjust the pressure regulator to the desired test pressure (e.g., 100 psi for standard API tests). Monitor the gauge carefully.
  5. Avvia test: Once the desired pressure is reached, immediately start a timer.
  6. Monitor Filtrate: Observe the volume of filtrate collected in the graduated cylinder over the specified test duration (e.g., 30 minutes for standard API tests).
  7. Risultati record: At the end of the test duration, record the total volume of filtrate collected.
  8. Pressione di rilascio: Slowly close the CO2 cylinder valve, then carefully open the pressure relief valve on the regulator to depressurize the system. Ensure the gauge reads zero before proceeding.
  9. Smontare e pulire: Carefully disassemble the pressure cell. Remove the filter paper and fluid sample. Proceed to cleaning as described in the Maintenance section.

6. Manutenzione

Regular maintenance ensures the longevity and accuracy of your API Filter Press.

6.1. Pulizia

6.2. O-Ring Inspection and Replacement

6.3. Pressure Regulator Care

7. Risoluzione Dei Problemi

ProblemaPossibile causaSoluzione
Pressure loss during testLeaking O-rings; Loose connections; Damaged pressure cell.Inspect and replace O-rings. Tighten all connections. Inspect pressure cell for cracks or damage; replace if necessary.
No filtrate collectedClogged filter paper; Insufficient pressure; Blocked filtrate outlet.Ensure fresh filter paper is used. Verify pressure setting. Clean filtrate outlet.
Risultati incoerentiimproprio sample preparation; Varying pressure; Contaminated equipment.Ensure consistent sample preparation. Verify pressure regulator accuracy. Thoroughly clean all components before each test.
Il manometro non legge correttamenteGauge malfunction; Regulator issue.Contact OFITE support for calibration or replacement of the pressure regulator/gauge.

8. Specifiche

9. Garanzia e supporto

OFITE products are manufactured to high standards of quality and reliability. For specific warranty terms and conditions, please refer to the documentation provided with your purchase or visit the official OFITE websito.

For technical support, spare parts, or service inquiries, please contact OFITE directly:

Quando contatti l'assistenza, assicurati di avere con te il modello del tuo prodotto e ASIN (B09CVBKG5L) readily available.

Documenti correlati - API Filter Press with CO2 Pressure Assembly

Preview OFITE 140-60 Complete Half-Area Filter Press Instruction Manual
Instruction manual for the OFITE 140-60 Complete Half-Area Filter Press, detailing its components, operation, maintenance, and warranty information. Designed for low-pressure filtration analysis of drilling fluids.
Preview Manuale di istruzioni del filtro pressa OFITE HTHP - Modelli 170-00-3 e 170-01-3
This comprehensive instruction manual details the operation, maintenance, and safety procedures for the OFITE High Temperature High Pressure (HTHP) Filter Press, including models 170-00-3 (115 V) and 170-01-3 (220 V). It covers specifications, components, testing procedures, and warranty information for drilling fluid analysis.
Preview OFITE 4-Unit HTHP Filter Press Instruction Manual
Instruction manual for the OFITE 4-Unit High-Temperature High-Pressure (HTHP) Filter Press, detailing setup, operation, safety procedures, load cell option, software setup, calibration, driver installation, temperature controller usage, and warranty information.
Preview Manuale di istruzioni del filtro pressa OFITE HTHP con celle filettate
Manuale di istruzioni per filtropressa ad alta temperatura e alta pressione (HTHP) OFITE con celle filettate (modelli 171-181-N, 171-181-1-N). Descrive dettagliatamente la valutazione di fluidi di perforazione, fanghi di cemento, ecc., a temperature e pressioni elevate, nel rispetto delle specifiche API. ...
Preview Filtropressa OFITE HTHP per prove su cemento - Manuale di istruzioni
Manuale di istruzioni per il filtro pressa OFITE HTHP (modello 170-182) progettato per valutare le caratteristiche di filtrazione dei fluidi di perforazione, delle fanghi di cemento e di altri fluidi in condizioni di alta temperatura e pressione.
Preview Filtropressa OFITE HTHP con celle filettate per prove su cemento - Manuale di istruzioni
Comprehensive instruction manual for the OFITE HTHP Filter Press with Threaded Cells (Model #171-192), designed for cement testing under high temperature and pressure conditions. Learn about specifications, operation, maintenance, and safety.